A STUDY OF THE SONIC RESONANCE OF THE FEMORAL PART OF HIP ENDOPROSTHESIS
Abstract
Mechanical resonance is the property of a mechanical system that responds to the oscillations coming from the outside of the
observed system. The response amplitude of the system is the highest when the frequency of the oscillations matches the system’s
natural frequency (its resonance frequency). It is known that resonance can cause swaying motions and even catastrophic
failure in improperly constructed structures. The phenomenon is known as resonance disaster. Sonic energy enters the system
through excitation and is dissipated through damping. Damping can be internal (within the material) or external (mounting of an
object). Six retrieved stems of hip endoprostheses were studied. For each of them sonograms were made, showing very distinctive
and narrow resonance curves with one major resonance peak followed by several higher harmonic peaks. Simulation of
endoprostheses standing waves was also performed resulting in the demonstration of various standing wave modes depending on
the observed frequency. Results clearly show that, due to the geometry and the used material, the observed endoprostheses have
very distinctive sonic resonance characteristics. The resonance is excited by the sound coming from outside (or inside) of the
human body. The energy of resonance movement of the endoprosthesis is dissipated through the endoprosthesis–bone interface.
A long-term exposure to the resonance oscillations might contribute to other causes of aseptic loosening of endoprostheses.
References
loosening, not only a question of wear A review of different theories,
Acta Orthopaedica, 77 (2006) 2, 177–197, doi:10.1080/
17453670610045902
2 J. Gallo, S. B. Goodman, Y. T. Konttinen, M. Raska, Particle disease:
Biologic mechanisms of periprosthetic osteolysis in total hip
arthroplasty, Innate Immun, 19 (2013) 2, 213–224, doi:10.1177/
1753425912451779
3 N. Verdonschot, R. Huiskes, Cement debonding process of total hip
arthroplasty stems, Clin. Orthop. Relat. Res. 3 (1997) 336,297-307,
doi:10.1097/00003086-199703000-00038
4 R. A. Cooper, C. M. McAllister, L. S. Borden, T. W. Bauer, Polyethylene
debris-induced osteolysis and loosening in uncemented total
hip arthroplasty, A cause of late failure, J Arthroplasty, 7 (1992) 3,
285-90, doi:10.1016/0883-5403(92)90050-z
5 D. Bitar, J. Parvizi, Biological response to prosthetic debris, World J.
Orthop., 18(2015)2, 172–189, doi:10.5312/wjo.v6.i2.172
6 E. Sukur, Y. E. Akman, Y. Ozturkmen, F. Kucukdurmazc, Particle
Disease: A Current Review of the Biological Mechanisms in
Periprosthetic Osteolysis After Hip Arthroplasty, Open Orthop J., 10
(2016) 241–251, doi:10.2174/1874325001610010241
7 U. Chong, A. Desmond, A. Hansen, Analysis of bone-endoprosthesis
interface micromotion for cementless tibial endoprosthesis fixation
and the influence of loading conditions, Journal of biomechanics, 43
(2010) 1074–1080, doi:10.1016/j.jbiomech.2009.12.006
8 B. A. Behrens, C. J. Wirth, H. Windhagen, I. Nolte, A. Meyer-
Lindenberg, A. Bouguecha, Numerical investigations of stress shielding
in total hip endoprostheses, Proc. Inst. Mech. Eng. H, 222 (2008)
5, 593–600, doi:10.1243/09544119JEIM139
9 B. Cheruvu, I. Venkatarayappa, T. Goswami, Stress Shielding in Cemented
Hip Implants Assessed from Computed Tomography, Biom.
Jour. of Sci. Tech. Res., 18 (2019) 3, 13637–13641, doi:10.26717/
BJSTR.2019.18.003163
10 A. Fahlgren, M. P. Bostrom, X. Yang, L. Johansson, U. Edlund, F.
Agholme, P. Aspenberg, Fluid pressure and flow as a cause of bone
resorption, Acta Orthop. 81 (2010) 4, 508–516, doi:10.3109/
17453674.2010.504610
11 Y. Yan, J. Hu, H. Lu, W. Wang, Genetic susceptibility to total hip
arthroplasty failure: a case–control study on the influence of MMP 1
gene polymorphism, Diagn. Pathol., 9 (2014) 177–179, doi:10.1186/
s13000-014-0177-9
12 A. Del Buono, V. Denaro, N. Maffulli, Genetic susceptibility to aseptic
loosening following total hip arthroplasty: a systematic review,
British Medical Bulletin, 101 (2012) 1, 39–55, doi:10.1093/bmb/
ldr011
13 K. Avsec, M. Jenko, M. Conradi, A. Kocijan, A. Vesel, J. Kova~, M.
Godec, I. Beli~, B. [etina Bati~, ^. Donik, M. Goren{ek, B.
Kocjan~i~, D. Dolinar, Surface Properties of Retrieved Cementless
Femoral Hip Endoendoprostheses Produced from a Ti6Al7Nb Alloy,
Coatings, 8 (2019) 868, doi:10.3390/coatings9120868
14 I. Beli~, B. Klop~i~, A. Logar, M. Jenko, D. Dolinar, M. Goren{ek,
B. Kocjan~i~, Mechanical Resonance of Femoral Part of Hip
Endoprosthesis as Possible Cause of Aseptic Loosening, Proc. of the
27th International Conference on Materials and Technologies,
Portoro`, Slovenia, 2019
15 A. A. Shabana, Theory of Vibration: An Introduction (Mechanical
Engineering Series), Vol. 1, 2nd Ed., Springer, Berlin, p. 350, 1995
16 The Fundamentals of Modal Testing, Application Note 243 – 3,
Agilent Technologies, http://www.modalshop.com/techlibrary/Fundamentals%
20of%20Modal%20Testing.pdf, 20.12.2019