EFFECT OF STERILIZATION ON THE SURFACE PROPERTIES OF Ti6Al7Nb ALLOY FEMORAL STEMS

  • Klemen Avsec Department for Orthopaedic Surgery, University Medical Centre Ljubljana, Zaloška 9, 1000 Ljubljana, Slovenia
  • Marjetka Conradi Institute of Metals and Technology, Lepi pot 11, 1000 Ljubljana, Slovenia
  • Monika Jenko Institute of Metals and Technology, Lepi pot 11, 1000 Ljubljana, Slovenia
  • Boštjan Kocjančič Department for Orthopaedic Surgery, University Medical Centre Ljubljana, Zaloška 9, 1000 Ljubljana, Slovenia
  • Mojca Debeljak University Rehabilitation Institute, Republic of Slovenia, Linhartova 51, 1000 Ljubljana, Slovenia
  • Matevž Gorenšek MD Medicina, Bohoričeva 5, 1000 Ljubljana, Slovenia
  • Drago Dolinar Department for Orthopaedic Surgery, University Medical Centre Ljubljana, Zaloška 9, 1000 Ljubljana, Slovenia
Keywords: hip endoprostheses, stem, Ti6Al7Nb alloy, cleaning, steam autoclave sterilization, surface properties, wettability, roughness, XPS

Abstract

We investigated thirty retrieved femoral components (cement-less stems made from Ti6Al7Nb alloy) of prematurely failed hip endoprostheses, due to aseptic loosening, infection and low-grade infection and two new stems (after their expiry date) for comparison. The main aim was to find the impact of cleaning and sterilization procedures of new and retrieved stems on the surface properties, i.e., surface wetting, roughness and surface chemistry. Our hypothesis was that the effect of cleaning and sterilization should not affect the meaningful difference of surface properties of retrieved and new stems of hip cement-less endoprostheses. Clean, new and retrieved stems of hip endoprostheses were sterilized using classical steam autoclave AC, procedure protocol (121 °C, 1.25 bar, 20 min). Surface wettability was determined by measuring the static contact angle of a water drop, the roughness was measured by optical 3D system. The surface chemistry was analysed by X-ray Photoelectron Spectroscopy (XPS). In wettability results, the contact angle varied from 70.6° to 91°, and the surface roughness (Sa) varied from 4.2 µm to 4.7 µm for all thirty measured stems. We found that surface properties measured on three different stem positions (neck, middle and root) differ significantly, we can conclude that corundum (Al2O3) grit-blasted surface used to obtain the necessary roughness for better osteointegration is not homogeneous. The XPS results of new and retrieved implants that prematurely failed for different reasons showed minor differences of TiO2 thin oxide native film thickness: 6.2 nm for the new stem and 5.2 nm for retrieved (prematurely failed due to aseptic loosening, infection or low grade infection, consisting primarily of TiO2). The results confirm our hypothesis that the effect of cleaning and sterilization on surface properties of new and retrieved stems is negligible.

References

1 S. Kurtz, K. Ong, E. Lau, F. Mowat, M. Halpern, Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030, J. Bone Joint Surg. Am., 89 (2007) 780–785
2 M. J. Grimm, Standard Handbook of Biomedical Engineering; McGraw-Hill, New York, USA, 2004
3 L. C. Jones, L. D.Timmie Topoleski, A. K.Tsao, Biomaterials in orthopedic Implants, Edt Elizabeth Friis; Mechanical testing in Orthopedic Implants, 2017, WP, Elsevier, doi:10.1016/B978-¬0-08-¬100286-¬5.00002-0
4 D. Dolinar, M. Gorenšek, M Jenko, M. Godec, B. Šetina, Č. Donik, A. Kocijan, M. Debeljak, B. Kocjančič, Biomaterials in endo¬prosthetics, Mater. Tehnol., 52, (2018)1, 89–98, doi:10.17222/mit. 2017.196
5 B. Mavčič, S. Roškar, V. Antoič, Clinical outcomes of titanium alloy SL-Plus femoral stem (Zweymuller): 2,013 total hip arthroplasty cases with 25 years of follow-up, 3rd International Symposium on Biomaterials, October 16–17, 2019, Portorož, 3ISB Abstract booklet IMT
6 M. J. Yaszemski, D. J. Trantolo, K. U. Levandrowski, V. Hasirci, D. E. Altobelli, Biomaterials in Orthopedics; CRC Press, New York, USA, 2003
7 M. Merola, S. Affatato, Materials for Hip Prostheses: A Review of Wear and Loading Considerations, Materials, 12 (2019) 495
8 M. Jenko, M. Gorenšek, M. Godec, M. Hodnik, B. Šetina, Č. Donik, J. T. Grant, D. Dolinar, Surface chemistry and microstructure of metallic biomaterials for hip and knee endoprostheses, Appl. Surf. Sci., 427 (2018) 584–593 doi/10.1016/japsuc2017
9 D. V. Kilpadi, J. J. Weimer, J. E. Lemons, Effect of passivation and dry heat-sterilization on surface energy and topography of unalloyed titanium implants, Colloids Surfs, A 135 (1998) 89–101
10 J. Göske, W. Uter, H. U. Ulrich, W. Kachler, G. Zeiler, A. Schuh, Surface Characterization of Corundum Blasted Implants in Hip Arthroplasty, Microsc. Microanal. 2004, 18, 9–11
11 R. K. Alla, K. Ginjupalli, N. Upadhya, M. Shammas, R. K. Ravi, R.Sekhar, Surface Roughness of Implants: A Review, Trends Biomater. Artif. Organs., 25 (2011) 112–118
12 C. Y. Guo, J. P. Matinlinna, J. K. H.Tsoi, Residual Contaminations of Silicon-Based Glass, Alumina and Aluminum Grits on a Titanium Surface After Sandblasting, Silicon 2015, 1–8
13 D. Bitar, J. Parvizi, Biological response to prosthetic debris, World J. Orthop., 6 (2015), 172–189
14 K. Avsec, M. Jenko, M. Conradi, A. Vesel, M. Mozetič, B. Kocjančič, D. Dolinar, Effect of autoclave or plasma oxygen gaseous sterilization on surface properties of cementless TiAl7Nb hip endoprostheses, 3rd International Symposium on Biomaterials, October 16–17 2019, Portorož, 3ISB Abstract booklet IMT
15 K. Avsec, M. Jenko, M. Conradi, A. Kocijan, A. Vesel, J. Kovac, M. Godec, I. Belic, B. Š. Batic, C. Donik, M.Gorenšek , B. Kocjancic, D. Dolinar, Surface Properties of Retrieved Cementless Femoral Hip Endoprostheses Produced from a Ti6Al7Nb Alloy, Coatings, 9 (2019) 868; doi:10.3390/coatings9120868
16 D. Dolinar, M. Jenko, M. Godec, M. Gorenšek, K. Avsec, B. Pompe, A. Cör, B. Kocjančič, Particulate debris from corundum-blasted titanium alloy cementless hip endoprostheses effecting aseptic loosening, AAOS 2020 Annual meeting, March 24–28,2020, Orlando Florida, E-presentation (no.) 626
17 J. H. Park, R. Olivares-Navarrete, R. E. Baier, A. E. Meyer, R. Tannenbaum, B. D. Boyan, Z. Schwartz, Effect of cleaning and sterilization on surface properties and cell response of titanium alloy implant surface, Acta Biomater., 8 (2012)5, 1966–1975
18 H. H. Tuson, D. B. Weibel, Bacteria-surface interactions, Soft Matter., 14 (2013) 9, 4368–4380, doi:10.1039/C3SM27705D
19 W. A. Rutala, D. J. Weber; Disinfection and sterilization: An overview, American Journal of Infection Control, 41 (2013) 52–55
20 B. Kausemo, J. Lausmaa, Biomaterial and implant surface – on the role of cleanliness, contamination, and preparation procedures, J. Biomed. Mater. Res. A, 22 (1988), 145–158
21 M. Raphel, S. B. Holodniy, S. C. Goodman, C. Heilshorn, Multi¬functional coatings to simultaneously promote osseointegration and prevent infection of orthopaedic implants. Biomaterials, 84 (2016) 301–314
22 A. G. Gristina, Biomaterial-centered infection: microbial adhesion versus tissue integration. Science 237 (1987) 1588–1595
23 P. Cools, N. De Geyter , E. Vanderleyden , P. Dubruel, R. Morent, Surface Analysis of Titanium Cleaning and Activation Processes,Non-thermal Plasma Versus Other Techniques, Plasma Chem Plasma Process, 34 (2014), 917–932, doi 10.1007/s11090-¬014-¬9552-2
24 ASTM F1295-05.Standard Specification for Wrought Titanium-6 Aluminum-7 Niobium Alloy for Surgical Implant Applications. (UNS R56700), https://infostore.saiglobal.com/store/details.aspx/ details.aspx?ProductID=249286 (10.06.2017).
25 D. Briggs, J. T. Grant,. Surface Analysis by Auger and X-ray Photoelectron Spectroscopy Surface Spectra; IM Publications: Manchester, UK, 2003
26 W. D. Kenton, B. A. Childs, A. Carlson, A. Lori, J. F. La Vannier, D. F. Moulder, Handbook of Auger Electron Spectroscopy; Physical Electronics: Eden Prarie, MN, USA, 1995
27 J. F. Moulder, P. E. Sobol, Handbook of X-ray Photoelectron Spectroscopy; Physical Electronics:Eden Prarie, MN, USA, 1995
Published
2021-03-16
How to Cite
1.
Avsec K, Conradi M, Jenko M, Kocjančič B, Debeljak M, Gorenšek M, Dolinar D. EFFECT OF STERILIZATION ON THE SURFACE PROPERTIES OF Ti6Al7Nb ALLOY FEMORAL STEMS. MatTech [Internet]. 2021Mar.16 [cited 2024Dec.10];55(1):59-4. Available from: https://mater-tehnol.si/index.php/MatTech/article/view/93